Adults 18 to 55, any sex, with Healthy Volunteers. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
Accuracy of Masimo W1™ and Apple Watch Series 8 in Determining SpO2 (Oxygen Saturation) Values in Subjects During DesaturationPrimary· 1-6 hours
Accuracy will be determined by comparing Masimo W1™ SpO2 measurements and Apple Watch Series 8 SpO2 measurements with FDA-cleared pulse oximetry device SpO2 measurements and calculating the Arithmetic root mean square (ARMS) value. All values presented below for SpO2 (Bias, Precision and ARMS) are in units of %.
The pairwise samples consist of paired SpO2 timing measurements from both the reference sensor, Masimo W1, and the Apple Watch.
Clinical Bias
Group
Value
95% CI
Masimo W1™
0.4
Apple Watch Series 8
0.8
Precision
Group
Value
95% CI
Masimo W1™
1.9
Apple Watch Series 8
2.0
Arithmetic Root Mean Square (ARMS)
Group
Value
95% CI
Masimo W1™
2.0
Apple Watch Series 8
2.2
Successful Read Rate for Motion Conditions Under Slow Desaturation for Masimo W1™ and Apple Watch Series 8Primary· 1-6 hours
This analysis examines the successful read rates for motion conditions during slow desaturation events for the Masimo W1™ and Apple Watch Series 8. The motion condition consists of a combined value representing three types of simulated motion: drinking, scratching the nose, and checking the time.
Successful Read Rate = 1 - No Read Rate (Nf / Nspotcheck) x 100 (%) Nf = Number of Reads without Valid SpO2 Nspotcheck = Number of All Spot Checks
Group
Value
95% CI
Masimo W1™
100
Apple Watch Series 8
4.7
Detection Rate of Fast Desaturation Events Under Various Conditions for Masimo W1™ and Apple Watch Series 8Primary· 1-6 hours
Comparison of the detection rate of fast desaturation events under different conditions for the Masimo W1™ and Apple Watch Series 8. The first condition involves no motion, with the palm facing down and the Apple Watch in sleep mode, while the second condition involves no motion, with the palm facing sideways and manual spot checks performed on the Apple Watch. The goal is to evaluate the performance of both devices in detecting desaturation events under these varying circumstances.
Detection Rate = (Nt / Ndesat) x 100 (%) Nt = Number of Detected Event by Test Device
All subjects who are enr
Condition 1: No motion, palm facing down, Apple Watch in Sleep Mode
Group
Value
95% CI
Masimo W1™
100
Apple Watch Series 8
7.4
Condition 2: No motion, palm facing sideways, manual spot checks for Apple Watch
Group
Value
95% CI
Masimo W1™
100
Apple Watch Series 8
24.7
Sponsor's own description
This study is also designed to compare the performance of Masimo W1™ and Apple Watch Series 8 against FDA cleared hospital grade pulse oximetry technology under desaturation conditions.
Publications & conference data
No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.
NCT07504003 — Influence of Training Session Duration on Improvements in Physiological Resilience to Exercise
· NA
· recruiting
NCT07446400 — A Trial to Examine the Interaction of Repinatrabit With Ethinyl Estradiol/Norethindrone, Metformin,Carbamazepine, Rosuva
· Phase 1
· recruiting
NCT07495813 — A Study to See How RO7763505 Works and How Safe it is When Given to Healthy People and People With Stable Heart Disease
· Phase 1
· recruiting
NCT07499050 — A Study of Multiple Doses of Orally Administered RO7795081 in Otherwise Healthy Chinese Adult Participants With Obesity
· Phase 1
· recruiting
NCT07342114 — A Dose-Escalation Study of RO7875913 in Healthy Participants
· Phase 1
· recruiting
Other Masimo Corporation trials
Trials by the same sponsor.
NCT07223177 — Philips FAST SpO2 Technology With Masimo Sensors for SpO2 Monitoring in the Neonatal, Infant, and Pediatric Populations
· NA
· recruiting
NCT07223164 — Validation of the Masimo Irregular Heartbeat Detection Algorithm in Participants Without Cardiovascular Disease
· NA
· recruiting
NCT06334055 — Validation Study of Mean Arterial Pressure (MAP) Parameter of Masimo INVSENSOR00073
· NA
· completed
NCT06432881 — Investigation of Skin Pigmentation Effect on Performance of Masimo Pulse Oximetry (INSPIRE)
· NA
· recruiting
NCT06148623 — Clinical Performance Testing of Philips FAST SpO2 With Masimo Pulse Oximetry Sensors Across Skin Pigmentation
· NA
· completed
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by Masimo Corporation
Last refreshed: 9 April 2025
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT05968079.