18 and older, any sex, with Astigmatism or Ametropia. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
Least Squares Means Distance Visual Acuity With Study Lenses at Week 1Primary· Day 8 of each study lens type worn during the corresponding crossover period
Distance visual acuity (VA) was assessed with study lenses in place. VA was collected for each eye separately using logarithmic of the Minimum Angle of Resolution (logMAR) charts. LogMAR typically ranges from -0.3 (20/10 vision on the Snellen chart) to 1 (20/200 vision). A logMAR acuity of 0.0 corresponds to 20/20 Snellen (normal distance visual acuity), with a negative value denoting better than 20/20 visual acuity.
Group
Value
95% CI
PRECISION1 for Astigmatism (P1fA)
-0.11
± 0.005
MyDay Toric (MDT)
-0.10
± 0.005
Sponsor's own description
The primary purpose of this study is to demonstrate noninferiority in the visual acuity at distance when wearing PRECISION1™ for Astigmatism (P1fA) soft contact lenses compared to another commercially available, soft toric contact lens, MyDay® toric (MDT).
Publications & conference data
No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by Alcon Research
Last refreshed: 2 November 2023
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT05483127.