19 and older, any sex, with Bradycardia or Sinus Node Dysfunction. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
Ventricular Capture Threshold, 3 Months Unipolar or BipolarPrimary· 3 months
Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed u
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
0.66
± 0.14
Left Bundle Area Pacing
0.65
± 0.13
QRS Duration - Intrinsic ConductionSecondary· at index procedure prior to lead fixation
Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles.
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
145
± 21.21
Left Bundle Area Pacing
124.83
± 26.63
QRS Duration After Lead FixationSecondary· at index procedure following final lead fixation
Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
129.33
± 35.8
Left Bundle Area Pacing
129.8
± 10.83
QRS Duration- 3 MonthsSecondary· 3 months
Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, 3 months after final pacing lead fixation, indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
108.67
± 19.22
Left Bundle Area Pacing
123
± 46.26
Capture Threshold - 6 Months, Unipolar or BipolarSecondary· 6 months
Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
1.18
± 0.79
Left Bundle Area Pacing
0.93
± 0.08
Capture Threshold - 12 Months, Unipolar or BipolarSecondary· 12 months
Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker
Group
Value
95% CI
Selective or Non Selective His Bundle Pacing
1.31
± 0.61
Left Bundle Area Pacing
1.12
± 0
Sponsor's own description
This is a voluntary research study to find out which location in the heart a pacemaker wire is the most efficient for a patient's heart and for battery life. Patients who volunteer and are eligible for the study will be randomized to receive one of two positions for the wire to be screwed into, in addition to studying multiple positions in the heart during the pacemaker insertion. Enrolled patients will be in the study for 1 year. They will also have an Ultrasound of their heart performed to assess how the pacemaker wire is affecting their heart. Pacemakers are connected to the heart by wires that are screwed into the heart. The wires can be connected to the heart in different places, which can affect how well the heart beats over time. The typical position is at the tip of the heart. This position may cause the heart to beat inefficiently. Over time, this could lead to weakened heart muscle, irregular heart rhythm, and more hospitalizations. The heart has special muscle cells and fibers that carry electrical signals through and around the heart. An alternative spot to place the pacemaker wire is in an area where these special cells are grouped together (called the HIS bundle). The pacemaker wire can be connected to the heart at a location which may allow the heart to beat more efficiently when compared to putting the wire at traditional spots in the heart (called HIS bundle pacing). However, sometimes connecting the wire into the HIS bundle may cause the pacemaker battery to wear out faster. Physicians can also connect the pacemaker wired near the HIS bundle (called Left left Bundle bundle area pacing). The study physicians hope this will allow the heart to beat more efficiently without causing the battery to wear out faster. The study physicians would like to study how different wire positions change heart beat efficiency and how long the pacemaker battery lasts when the wires are placed in different locations. This study will connect the pacemaker wire at either the HIS Bundle or the left bundle area pacing, to see how effectively the heart pumps and how much battery is being used.
Publications & conference data
No peer-reviewed publications indexed yet for this trial.
NCT07209852 — Safety, Performance, and Clinical Benefit of Pacing the Left Bundle Branch Area - Post-Market Clinical Follow-up
· recruiting
NCT07254013 — Left Bundle Branch Area Pacing (LBBAP) Monitoring Assistant: Evaluation of Portable ECG Solution
· NA
· recruiting
NCT07166614 — Dexmedetomidine vs Propofol in High-Risk ERCP Patients
· NA
· recruiting
NCT07585799 — Permanent Conduction System Pacing Versus Atrioventricular Management for Patients With Wide QRS and Patent AV Conductio
· NA
· recruiting
NCT07217392 — Left Bundle Branch Area Pacing (LBBAP) PMCF Study
· NA
· recruiting
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by David Haines, MD
Last refreshed: 17 August 2022
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT04093414.