NCT04093414: LEFTBASH

Left Bundle Area Versus Selective His Bundle Pacing

Terminated NA Results posted Last updated 17 August 2022
What this trial tests

NA trial testing Select Secure pacing lead in Bradycardia in 9 participants. Terminated before completion.

Timeline
7 October 2019
Primary endpoint
14 February 2022
22 February 2022

Quick facts

Lead sponsorDavid Haines, MD
PhaseNA
StatusTerminated
Study typeINTERVENTIONAL
Allocationrandomized
Designparallel
Maskingnone
Primary purposetreatment
Enrollment9
Start date7 October 2019
Primary completion14 February 2022
Estimated completion22 February 2022
Sites2 locations across United States

Drugs / interventions tested

Conditions studied

Sponsor

David Haines, MD

Who can join

19 and older, any sex, with Bradycardia or Sinus Node Dysfunction. Patients with the condition only — healthy volunteers not accepted.

Results — posted to ClinicalTrials.gov

Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.

Ventricular Capture Threshold, 3 Months Unipolar or Bipolar Primary · 3 months

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker settings are programmed u

GroupValue95% CI
Selective or Non Selective His Bundle Pacing0.66± 0.14
Left Bundle Area Pacing0.65± 0.13
QRS Duration - Intrinsic Conduction Secondary · at index procedure prior to lead fixation

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles.

GroupValue95% CI
Selective or Non Selective His Bundle Pacing145± 21.21
Left Bundle Area Pacing124.83± 26.63
QRS Duration After Lead Fixation Secondary · at index procedure following final lead fixation

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, with intrinsic conduction (prior to implantation of the pacemaker), indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.

GroupValue95% CI
Selective or Non Selective His Bundle Pacing129.33± 35.8
Left Bundle Area Pacing129.8± 10.83
QRS Duration- 3 Months Secondary · 3 months

Duration (time in milliseconds) of the QRS wave complex interval measured from the end of the PR interval to the end of the S wave measured on a 12-lead electrocardiogram, 3 months after final pacing lead fixation, indicating the length of time required for the electrical depolarization of the right and left ventricles of the heart and contraction of the large ventricular muscles. A normal duration is between 80-100 milliseconds. A QRS duration of greater than 120 milliseconds is considered abnormal.

GroupValue95% CI
Selective or Non Selective His Bundle Pacing108.67± 19.22
Left Bundle Area Pacing123± 46.26
Capture Threshold - 6 Months, Unipolar or Bipolar Secondary · 6 months

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker

GroupValue95% CI
Selective or Non Selective His Bundle Pacing1.18± 0.79
Left Bundle Area Pacing0.93± 0.08
Capture Threshold - 12 Months, Unipolar or Bipolar Secondary · 12 months

Ventricular capture threshold is the minimum amplitude of electrical signal from the pacemaker that consistently results in capture of the ventricular myocardium (normal contraction of the ventricle after electrical depolarization) with a 1.0 millisecond pulse width setting in unipolar or Bipolar output modes, whichever value is lower, measured in volts. During pacemaker placement, the ideal ventricular capture threshold is determined by delivering a series of pulses starting at 0.75 volts and increasing by 0.125 V with each trial until consistent ventricular contraction is achieved. Pacemaker

GroupValue95% CI
Selective or Non Selective His Bundle Pacing1.31± 0.61
Left Bundle Area Pacing1.12± 0

Sponsor's own description

This is a voluntary research study to find out which location in the heart a pacemaker wire is the most efficient for a patient's heart and for battery life. Patients who volunteer and are eligible for the study will be randomized to receive one of two positions for the wire to be screwed into, in addition to studying multiple positions in the heart during the pacemaker insertion. Enrolled patients will be in the study for 1 year. They will also have an Ultrasound of their heart performed to assess how the pacemaker wire is affecting their heart. Pacemakers are connected to the heart by wires that are screwed into the heart. The wires can be connected to the heart in different places, which can affect how well the heart beats over time. The typical position is at the tip of the heart. This position may cause the heart to beat inefficiently. Over time, this could lead to weakened heart muscle, irregular heart rhythm, and more hospitalizations. The heart has special muscle cells and fibers that carry electrical signals through and around the heart. An alternative spot to place the pacemaker wire is in an area where these special cells are grouped together (called the HIS bundle). The pacemaker wire can be connected to the heart at a location which may allow the heart to beat more efficiently when compared to putting the wire at traditional spots in the heart (called HIS bundle pacing). However, sometimes connecting the wire into the HIS bundle may cause the pacemaker battery to wear out faster. Physicians can also connect the pacemaker wired near the HIS bundle (called Left left Bundle bundle area pacing). The study physicians hope this will allow the heart to beat more efficiently without causing the battery to wear out faster. The study physicians would like to study how different wire positions change heart beat efficiency and how long the pacemaker battery lasts when the wires are placed in different locations. This study will connect the pacemaker wire at either the HIS Bundle or the left bundle area pacing, to see how effectively the heart pumps and how much battery is being used.

Publications & conference data

No peer-reviewed publications indexed yet for this trial.

Verify or expand the search:

Other recruiting trials for Bradycardia

Currently open trials in the same condition.

Verify against primary sources

Data sources for this page

Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT04093414.

Primary sources · FDA · ClinicalTrials.gov · EMA · SEC EDGAR · ChEMBL · Wikidata · full sourcing