NCT06896734: SEERS

mobiCARE™ ECG Monitoring System

Completed Results posted Last updated 9 May 2025
What this trial tests

trial testing Evaluation of the ECG signal in the nominal location and in three alternate positions in ECG Monitoring in 49 participants. Completed in 16 March 2025.

Timeline
1 January 2025
Primary endpoint
5 February 2025
16 March 2025

Quick facts

Lead sponsorSeers Technology Co., Ltd.
StatusCompleted
Study typeOBSERVATIONAL
Enrollment49
Start date1 January 2025
Primary completion5 February 2025
Estimated completion16 March 2025
Sites1 location across United States

Drugs / interventions tested

Conditions studied

Sponsor

Seers Technology Co., Ltd.

Who can join

18 and older, any sex, with ECG Monitoring. Patients with the condition only — healthy volunteers not accepted.

Results — posted to ClinicalTrials.gov

Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.

Number of Subjects in Which the ECG Signal is Interpretable at the Nominal Location and up to 3 Alternate Positions Primary · 2 minutes while the patient ambulates

Evaluation of the ECG signal at the nominal location and in up to three (3) additional alternate positions (AP) in each subject to determine if the signal is interpretable. The term "interpretable" is defined as follows: An independent cardiologist reader is able to identify the ECG rhythm in each location when compared to the ECG rhythm measured in the nominal position of the mobiCARE™ Cardiac Monitoring System described in the Instructions for Use. ECG rhythm data will be collected for up to two (2) minutes while the patient ambulates. ECG rhythm data were collected in each location during

GroupValue95% CI
Nominal Location49
Alternate Position 149
Alternate Position 249
Alternate Position 349
Presence of Artifact Graded Secondary · 2 minutes

Presence of artifact graded as 1=none, 2=minor (cardiac rhythm is interpretable), or 3=major (cardiac rhythm is not interpretable by the independent cardiologist).

GroupValue95% CI
Nominal Location35
Alternate Position 121
Alternate Position 228
Alternate Position 331
Nominal Location9
Alternate Position 114
Alternate Position 211
Alternate Position 311
Nominal Location5
Alternate Position 114
Alternate Position 210
Alternate Position 37
Nominal Location0
Alternate Position 10
Alternate Position 20
Alternate Position 30
Evaluation of the ECG Morphology Measurements at the Nominal Position and up to 3 Alternate Positions Secondary · 2 Minutes

The number of measurable positions for QRS amplitude and QRS duration at the nominal position and up to 3 alternate positions.

GroupValue95% CI
Nominal Position49
Alternate Position 149
Alternate Position 249
Alternate Position 349
Human Factors Evaluation Measures Secondary · 2 Minutes

Usability Performance Evaluation for Health Care Professionals and Subjects

How easy/difficult for the HCP to place device/electrodes in the Nominal Location using the IFU?
GroupValue95% CI
Limited Human Factors Evaluation Measures3.94± 0.92
Could the HCP read the ECG signal on app to confirm Lead II/clear signal after the device placed?
GroupValue95% CI
Limited Human Factors Evaluation Measures4.43± 0.74
How easy/difficult for subject to remove the device/electrodes from Alternate location 3?
GroupValue95% CI
Limited Human Factors Evaluation Measures4.1± 1.01
How easy or difficult was it for the subject to remove the used electrodes from the mobiCARE™?
GroupValue95% CI
Limited Human Factors Evaluation Measures3.67± 1.13
How easy/difficult for subject to replace electrodes with new electrodes on the mobiCARE™ device?
GroupValue95% CI
Limited Human Factors Evaluation Measures3.69± 1.29
How easy or difficult was it for the subject to locate the markings for the nominal position?
GroupValue95% CI
Limited Human Factors Evaluation Measures3.49± 1.23
How easy/difficult for subject to replace new mobiCARE™ device (electrodes) in the Nominal position
GroupValue95% CI
Limited Human Factors Evaluation Measures3.78± 1.23
The Number of Measurable Positions for P-waves Secondary · 2 minutes

Informational data collected to evaluate interpretability based on P-wave amplitude

GroupValue95% CI
Nominal Position48
Alternate Position 148
Alternate Position 248
Alternate Position 348

Sponsor's own description

Evaluation of the mobiCARE™ ECG Monitoring System Electrode Placement Positioning Limits in a Prospective, Non-Randomized, Single-Center U.S. Study

Publications & conference data

No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.

Verify or expand the search: