Adults 1 Minute to 12, any sex, with Measure Accuracy of Respiratory Rate. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
Accuracy of Device Under Test for the Measurement of Respiratory Rate, Measured in Breath/MinPrimary· Every 60 seconds, over a total duration of maximum 60 minutes
Arms or rms error, given in breath per minute, is the calculation of root mean square between Respiratory rate of the subject device and the manually scored reference Respiratory Rate EtCO2 waveform, for all stable respiratory periods. It is a way of averaging the absolute values of errors over the full measurement range and is a similar measurement to a "rms error". It indicates how far (accurate) the provided value is from the reference used. A passing result required an Arms of ≤3 breaths/min.
Group
Value
95% CI
Pediatrics Population
2.9
Sponsor's own description
The purpose of this study is to conduct a Respiratory Rate accuracy validation on pediatrics population comparing the Gabi SmartCare Gabi Band to the Reference, an FDA cleared End Tidal Carbon Dioxide monitor (GE Datex- Ohmeda) by manually scoring the collected waveform for data analysis.
Publications & conference data
No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by Gabi SmartCare
Last refreshed: 26 November 2024
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT05856422.