Adults 18 to 39, any sex, with Visual Acuity. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
High Luminance High Contrast Monocular Visual Acuity (LogMAR)Primary· 1-Week Follow-up
Visual Acuity was collected for each subject eye at distance (4 meter) using a Logarithm of Minimal Angle of Resolution (logMAR) visual acuity scale. This was evaluated under high luminance and high contrast conditions (HLHC) at 4 meters from Early Treatment Diabetic Retinopathy Study (ETDRS) charts at the 1-week follow-up visit. A value of 0.0 logMAR is equivalent to 20/20 Snellen Visual Acuity. Lower values of visual acuity indicate better vision. The average logMAR Visual Acuity was reported for each lens type.
Group
Value
95% CI
Test (AO1DfA)
-0.126
± 0.0798
Control (AO1D)
-0.028
± 0.1123
CLUE Vision ScoreSecondary· 1-Week Follow-up
CLUE Vision Score was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The average CLUE vision score for each lens type was report
Group
Value
95% CI
Test (AO1DfA)
69.39
± 20.151
Control (AO1D)
41.49
± 28.139
High Luminance High Contrast Monocular Visual Acuity (LogMAR)Secondary· 1-Week Follow-up
Visual Acuity was collected for each subject eye at distance (4 meter) using a Logarithm of Minimal Angle of Resolution (logMAR) visual acuity scale. This was evaluated under high luminance and high contrast conditions (HLHC) at 4 meters from Early Treatment Diabetic Retinopathy Study (ETDRS) charts at the 1-week follow-up visit. A value of 0.0 logMAR is equivalent to 20/20 Snellen Visual Acuity. Lower values of visual acuity indicate better vision. The average logMAR Visual Acuity was reported for each lens type. Data summaries for the Control lens are reported for the first primary measure o
This will be a 5-visit, randomized, partially single-masked, bilateral wear, dispensing, 2-treatment × 2-period crossover study with spectacle-wear washout and wash-in periods.
Publications & conference data
No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.
NCT07443215 — Evaluation of Effect of Scatter on Visual Performance
· NA
· recruiting
NCT07135193 — Evaluation of the Protective Effects of Cordyceps Cicadae on Visual Function and Anti-Fatigue Performance in Esports Ath
· NA
· active not recruiting
NCT07085533 — Natural History Study of Inherited Retinal Diseases
· recruiting
NCT07135180 — Exploring the Eye Care Benefits of Cordyceps Cicadae Across Different Age Groups
· NA
· active not recruiting
NCT04632901 — A New Parameter for the Assessment of Distance Visual Capacity: the Critical Visual Acuity (CVA)
· active not recruiting
Other Johnson & Johnson Vision Care, Inc. trials
Trials by the same sponsor.
NCT07443215 — Evaluation of Effect of Scatter on Visual Performance
· NA
· recruiting
NCT07174596 — Evaluation of Approved Contact Lenses
· NA
· completed
NCT07076160 — Pilot Dispensing Study of Two Marketed Daily Disposable Multifocal Contact Lenses
· NA
· completed
NCT06967129 — Evaluation of Daily Disposable Toric Soft Lenses Manufactured With an Alternative Hydration Process
· NA
· completed
NCT06864858 — Clinical Evaluation of Two Cosmetic Contact Lenses
· NA
· completed
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by Johnson & Johnson Vision Care, Inc.
Last refreshed: 27 April 2023
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT05056870.