Intramuscular Heating Rates of a Chattanooga Intelect XT Therapeutic Ultrasound

Completed NA Results posted Last updated 27 May 2021
What this trial tests

NA trial testing Therapeutic Ultrasound in Ultrasound Therapy; Complications in 25 participants. Completed in 11 May 2020.

Timeline
5 February 2019
Primary endpoint
15 April 2020
11 May 2020

Quick facts

Lead sponsorNorth Dakota State University
PhaseNA
StatusCompleted
Study typeINTERVENTIONAL
Allocationna
Designsingle group
Maskingnone
Primary purposebasic science
Enrollment25
Start date5 February 2019
Primary completion15 April 2020
Estimated completion11 May 2020
Sites1 location across United States

Drugs / interventions tested

Conditions studied

Sponsor

North Dakota State University

Who can join

Adults 18 to 35, any sex, with Ultrasound Therapy; Complications. Patients with the condition only — healthy volunteers not accepted.

Results — posted to ClinicalTrials.gov

Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.

Heating Rate at 1 cm Depth Primary · Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 6 minutes)

A thermocouple was inserted at the 1 cm depth to determine the tissue temperatures.

GroupValue95% CI
Therapeutic Ultrasound Treatment6.49± 1.24
Heating Rate at 1.75 cm Depth Primary · Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 8 minutes)

A thermocouple was inserted at the 1.75 cm depth to determine the tissue temperatures.

GroupValue95% CI
Therapeutic Ultrasound Treatment5.50± 1.34
Heating Rate at 2.5 cm Depth Primary · Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which did not occur in the 15 minute treatment)

A thermocouple was inserted at the 2.5 cm depth to determine the tissue temperatures.

GroupValue95% CI
Therapeutic Ultrasound Treatment3.07± 1.53

Sponsor's own description

The purpose of this study is to determine the rate of tissue temperature increase in the triceps surae during an ultrasound treatment using the Chattanooga Intelect Legend XT. The following parameters will be used: 3 MHz frequency, 1.0 W/cm2 intensity, 15 minute treatment time. Thermocouples will be inserted at 1.0, 1.75, and 2.5 cm depths to determine the heating rates. Most modality textbook recommendations for therapeutic ultrasound are based on studies performed with the Omnisound ultrasound machine. Previous research has shown variations in power output across manufacturers, producing differences in tissue temperature increases. No information has been published regarding the heating rates of the Chattanooga Intelect Legend XT machine for ultrasound treatments. By determining the heating rates for this machine, clinicians will be better able to utilize the Chattanooga Intelect Legend XT to help reach therapeutic goals.

Publications & conference data

No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.

Verify or expand the search:

Other trials of Therapeutic Ultrasound

Trials testing the same drug.

Other recruiting trials for Ultrasound Therapy; Complications

Currently open trials in the same condition.

Other North Dakota State University trials

Trials by the same sponsor.

Verify against primary sources

Data sources for this page

Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT03816709.

Primary sources · FDA · ClinicalTrials.gov · EMA · SEC EDGAR · ChEMBL · Wikidata · full sourcing