visuALL Field Analyzer (vFA) Compared to Standard Humphrey Automated Perimetry

Completed NA Results posted Last updated 29 June 2022
What this trial tests

NA trial testing Standard Automatic Perimetry Humphrey Field Analyzer in Glaucoma, Open-Angle in 72 participants. Completed in 20 October 2020.

Timeline
8 January 2019
Primary endpoint
20 October 2020
20 October 2020

Quick facts

Lead sponsorWills Eye
PhaseNA
StatusCompleted
Study typeINTERVENTIONAL
Allocationnon randomized
Designparallel
Maskingnone
Primary purposediagnostic
Enrollment72
Start date8 January 2019
Primary completion20 October 2020
Estimated completion20 October 2020
Sites1 location across United States

Drugs / interventions tested

Conditions studied

Sponsor

Wills Eye — full company profile →

Who can join

Adults 21 to 80, any sex, with Glaucoma, Open-Angle. Patients with the condition only — healthy volunteers not accepted.

Results — posted to ClinicalTrials.gov

Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.

Mean Retinal Sensitivity From Two Machines Primary · three hours

Mean (Average) retinal sensitivity to light as measured in decibels in multiple spots across the central and peripheral retina (entire visual field). Higher numbers translate to dimmer light which when detected means better vision.

Global field from Humphrey Visual Field
GroupValue95% CI
Healthy Controls29.9± 1.4
Glaucoma Group24.8± 3.9
Global field from VisuALL
GroupValue95% CI
Healthy Controls28.7± 1.3
Glaucoma Group22.9± 4.9
Inferior nasal quadrant from Humphrey Visual Field
GroupValue95% CI
Healthy Controls30.3± 1.3
Glaucoma Group24.8± 5.8
Inferior nasal quadrant from VisuALL
GroupValue95% CI
Healthy Controls28.7± 1.6
Glaucoma Group23.3± 6.9
Superior hemifield from Humphrey Visual Field
GroupValue95% CI
Healthy Controls29.5± 1.5
Glaucoma Group23.7± 5.0
Superior hemifield from VisuALL
GroupValue95% CI
Healthy Controls28.6± 1.6
Glaucoma Group21.8± 6.5
Global Correlation (Agreement Between Two Machines in the Overall Visual Field) Primary · three hours

Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the central and peripheral retina (entire visual field). This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the visual field (overall), between two testing machines.

GroupValue95% CI
Healthy Controls0.50 – 1
Glaucoma Group0.80 – 1
Inferior Nasal Quadrant Correlation (Agreement Between Two Machines in the Inferior Nasal Visual Field) Secondary · three hours

Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the inferior nasal visual field. This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the inferior (lower) nasal portion of the visual field, between two testing machines.

GroupValue95% CI
Healthy Controls0.30 – 1
Glaucoma Group0.80 – 1
Superior Hemifield Correlation (Agreement Between Two Machines in the Superior Hemifield Visual Field Secondary · three hours

Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the superior (upper) hemifield portion of the visual field. This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the visual field in the superior hemifield, between two testing machines.

GroupValue95% CI
Healthy Controls0.40 – 1
Glaucoma Group0.60 – 1

Sponsor's own description

To determine age-adjusted reference values of the visuALL Field Analyzer (vFA) retinal sensitivity and to assess the repeatability of the measured values and compare them with those measured by a Standard Automatic Perimetry (SAP).

Publications & conference data

No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.

Verify or expand the search:

Other trials of Standard Automatic Perimetry Humphrey Field Analyzer

Trials testing the same drug.

Other recruiting trials for Glaucoma, Open-Angle

Currently open trials in the same condition.

Other Wills Eye trials

Trials by the same sponsor.

Verify against primary sources

Data sources for this page

Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT03804684.

Primary sources · FDA · ClinicalTrials.gov · EMA · SEC EDGAR · ChEMBL · Wikidata · full sourcing