5 and older, any sex, with Fever. Patients with the condition only — healthy volunteers not accepted.
Results — posted to ClinicalTrials.gov
Per-arm endpoint measurements with 95% confidence intervals where reported. Source: trial results section.
Clinical Agreement Through Clinical Bias EvaluationPrimary· one day
Agreement through clinical bias evaluation was calculated as the Mean Bias ± Standard Deviation (SD) by subtracting the first output measurement recorded with the welloStationX thermometer (DUT) from the corresponding output measurement recorded with the Welch Allyn SureTemp oral thermometer for each subject, and then calculating the Mean Bias ± SD of the biases.
Group
Value
95% CI
Temperature Measurement
0.04
± 0.42
Clinical RepeatabilitySecondary· one day
Clinical Repeatability was calculated using the pooled standard deviations of the triplicate measurements recorded by the welloStationX.
Group
Value
95% CI
Temperature Measurement
0.062
Sponsor's own description
This study evaluates the clinical accuracy of the welloStationX, an automated non-touch self-service electronic thermometer using an infrared sensor of the surface of the forehead to screen for fever.
Publications & conference data
No peer-reviewed publications indexed yet for this trial. Completed trials usually publish results within 12-18 months.
Drug + disease cross-links: matched in real time against Drug Landscape's normalised drug + company + condition tables
Sponsor: as reported to ClinicalTrials.gov by Wello Inc.
Last refreshed: 7 July 2021
Drug Landscape aggregates and links these public records for informational use only. Always verify against the primary source before clinical or regulatory decisions. Canonical URL: https://druglandscape.com/trial/NCT03758157.